Gas station without pumps

2012 August 9

Medical Instrumentation, Chapter 6

As I mentioned in Medical Instrumentation, first 5 chapters, for the past few days, I’ve been slogging through one of the potential text books for the circuits course: Medical Instrumentation: Application and Design, 4th Edition. John G. Webster. Publisher: Wiley, 2009. # ISBN-10: 0471676004; # ISBN-13: 978-0471676003.
Here are my notes on Chapter 6, which may be the most relevant chapter:
BIOPOTENTIAL AMPLIFIERS                            241
Michael R. Neuman

6.1         Basic Requirements   241
2-page intro about why amplifiers are needed and why input impedance needs to be at least 100 MΩ and gain around 1000. Explains need for bipolar signals and for fixed, calibrated gain.

6.2         The Electrocardiograph   243
Explains how the 10 wires (4 limb wires and 6 chest wires) produce the 12 standard signals of the standard 12-lead EKG. Talks about the heart as a varying electrical dipole, and how the 12 leads provide 60 degree spacing of vectors in 2 different planes.

Defines “Wilson’s central terminal” and explains how it can be used to get differential signals to get 6 directions out of the 4 limb wires.

Gives block diagram of a typical EKG.

Discusses the need for protection circuitry and ability for EKG to recover quickly from defibrillation shocks.

This section looks very valuable for our EKG lab.

6.3         Problems Frequently Encountered   254

Discusses many of the problems of EKG design (and, to a lesser extent, EEG design) caused by large transients, ground loops, and capacitive couple of 60Hz noise.  Again, an excellent section for students to read for the EKG lab.

6.4         Transient Protection    264

Parallel Si diodes or back-to-back Zener diodes to short out voltages that are too high.  (Neon bulbs are also mentioned, but their breakdown voltages seem a little high for EKGs.)

6.5         Common-Mode and Other Interference-Reduction Circuits   266

Describes shielding and driven-right-leg system. The driven-right-leg system has a lower effective resistance for the ground electrode than connecting the leg to ground, by feeding back the common-mode signal.  Because the ina126p instrumentation amplifier does not provide a common-mode output, this approach is not available to us.

6.6         Amplifiers for Other Biopotential Signals   269

Figure 6.16 gives frequency and voltage ranges for EOG, EEG, ECG, EMG, and AAP (axon action potential) amplifiers. EMG amplifiers need higher frequency response than ECG (but not as much low frequency response).

Glass micropipette electrodes may need negative-input-capacitance (positive feedback) amplifiers.

EEGs, because of small differential signals and high gain may need very high CMRR (example give needs 112 dB).

6.7         Example of a Biopotential Preamplifier   274

Gives the circuit for an EKG preamp, using op amps. Does not mention the difficulty of matching resistors accurately.  Does discuss saturation caused by DC offsets (hence the need for small gain), hi-pass filtering after the differential amplifier, and low-pass filter in the final stage.  Also has a shorting switch for recentering the stage after the high-pass filter after a transient that moves it out of the linear region.

The discussion of the design could be useful for our EKG lab, though the details will be different for the student design, since they’ll use a single-chip instrumentation amp and 1 resistor instead of 3 op amps and 7 resistors.

They also cite a 2-electrode design

Bootstrapped two-electrode biosignal amplifier:
Dobromir Petkov Dobrev, Tatyana Neycheva and Nikolay Mudrov
Medical and Biological Engineering and Computing
Volume 46, Number 6 (2008), 613-619, DOI: 10.1007/s11517-008-0312-4

that looks interesting.  It sticks a cross-coupled feedback stage in front of an instrumentation amp, to get a low common-mode impedance with a high differential mode impedance.

Dobrev et al’s two-electrode amplifier that uses feedback into the differential electrodes to reduce the common-mode input impedance. I’ve not tested the circuit myself, but I like the idea of a 2-electrode EKG. This circuit won’t fit on the instrumentation amp protoboard, though, so I may want to redesign that board if we want to encourage designs like this one.  The op amps they chose have somewhat better input bias current and common-mode rejection, but lower gain-bandwidth product than the MCP6002 I’ve been using (we don’t need high bandwidth for an EKG).  It’s going to take me a while to understand this circuit well enough that I can explain every component.  I may want to build and test it also, to see if it really does work as claimed without a 3rd ground electrode.  (Schematic copied from article by Dobrev et al.)

Note: Dobrev et al. claim that the standard ECG bandwidth is 0.01-100Hz, but it looks to me like their reference voltage low-pass feedback may compromise the low end a bit. I may have to find a circuit simulator that I can run on my Mac that can handle these components (I’ve been using the web tool CircuitLab, which has a very limited library of components)

I’m a bit dubious of the need for frequencies below about 0.1 Hz. Heartbeats only go down to about 0.46 Hz for a healthy human, though rates as low as 0.16Hz have been reported in the medical literature (in a patient who died within 12 hours). Still, one doesn’t need the DC component of that waveform, and the interesting parts of the waveform are of much shorter duration. Note: bradycardia is often diagnosed at <60bpm (1Hz) and extreme bradycardia is diagnosed at <40bpm (0.67 Hz), though my resting rate is 42–52bpm (0.7–0.87 Hz)—a low heart rate that I believe I inherited from my father.

6.8         Other Biopotential Signal Processors   275

The described technique for the cardiotachometer seems quaint, using counters and registers, even though they say that “software calculates v_o”—the design obviously dates from the days when A-to-D conversion was expensive (say 25 years ago).  Nowadays, the whole thing would be done in software, with just enough analog amplification and filtering to make the signal suitable for input to an A-to-D input of a microprocessor.

The digital integration for the EMG makes more sense, though I wonder whether RMS makes more sense than mean absolute value.

Signal averaging is introduced for evoked potentials (examples given for EEG or ERG signals, as well as fetal EKG).  Other ways of separating fetal EKG from maternal EKG are discussed.

Vector cardiographs are mentioned, but rather dismissively.

6.9         Cardiac Monitors   282

They say that cardiac monitors typically have a bandwith of 0.67–40Hz, rather than the ?–150 Hz of diagnostic EKGs described in Section 6.2, the 0.01–250 Hz of Table 1.1, or the 0.01–100Hz mentioned by Dobrev.  It seems that there is quite a range of different ideas about what part of the waveform needs to be preserved.

Discussion includes the advantages of implantable monitors (like the Medtronics Reveal implantable cardiac monitor).

Block diagrams for lead fall-off alarms are also included.

6.10       Biotelemetry   287

Very brief mention of telemetry, and even briefer mention of modern methods (like ZigBee and BlueTooth).

Problems   288

References   291

Bottom line

I still don’t think that we can make Medical Instrumentation our main text book, but I like a lot of Chapter 6 (particularly when combined with the better parts of Chapters 4 and 5).  I think we’re going to have to extract and boil down the most useful material, though, as there is a bit too much and the useful details are too scattered to have the students go on a treasure hunt for it in the limited time that library reserves circulate for.  Writing the handout for the EKG lab is going to be a massive undertaking—almost a book chapter on how EKGs work.

About these ads

6 Comments »

  1. [...] Perhaps trying a 2-electrode, rather than 3-electrode, EKG (as described by Dobrev et al, the design I copied and cited in the post Medical Instrumentation Chapter 6). [...]

    Pingback by EKG blinky boards arrived « Gas station without pumps — 2012 August 13 @ 22:20 | Reply

  2. [...] also want to try Dobrev’s 2-wire EKG circuit (which I referred to in an earlier post), but that needs more op amps than the Rev 2.1 protoboard has room for: 5 op amps plus the [...]

    Pingback by Instrumentation amp prototyping boards arrived « Gas station without pumps — 2012 August 18 @ 16:12 | Reply

  3. [...] I mentioned in Medical Instrumentation, first 5 chapters and Medical Instrumentation, Chapter 6, I’ve been slogging through one of the potential text books for the circuits course: Medical [...]

    Pingback by Medical Instrumentation, Chapters 7 and 8 « Gas station without pumps — 2012 August 19 @ 13:09 | Reply

  4. [...] I mentioned in Medical Instrumentation, first 5 chapters, Medical Instrumentation, Chapter 6,  and Medical Instrumentation, Chapters 7 and 8, I’ve been slogging through one of the [...]

    Pingback by Medical Instrumentation, Chapters 9–14 « Gas station without pumps — 2012 August 19 @ 22:07 | Reply

  5. [...] Medical Instrumentation, Chapter 6 [...]

    Pingback by 2012 in review « Gas station without pumps — 2012 December 31 @ 11:18 | Reply

  6. […] Medical Instrumentation, Chapter 6 […]

    Pingback by Blogoversary 3 | Gas station without pumps — 2013 June 1 @ 20:00 | Reply


RSS feed for comments on this post. TrackBack URI

Leave a Reply

Fill in your details below or click an icon to log in:

WordPress.com Logo

You are commenting using your WordPress.com account. Log Out / Change )

Twitter picture

You are commenting using your Twitter account. Log Out / Change )

Facebook photo

You are commenting using your Facebook account. Log Out / Change )

Google+ photo

You are commenting using your Google+ account. Log Out / Change )

Connecting to %s

The Rubric Theme. Create a free website or blog at WordPress.com.

Follow

Get every new post delivered to your Inbox.

Join 274 other followers

%d bloggers like this: