Common-mode noise in EKG

Filed under: Circuits course — gasstationwithoutpumps @ 18:00
Tags: , , , , , , , ,

In 2-op-amp instrumentation amp, I said

I’ll have to make some measurements later this week to see how large the common-mode noise on the EKG signals really is. Of course, it is likely to be highly variable, depending on the electrodes and the wiring to them, but ball-park estimates would be useful.

If the AC common-mode voltage is ±200mV and we have worst-case resistor values, then we would have a ±8mV common-mode output from the instrumentation amp. With the lowest differential amplification (6.6 at Rgain=∞), a 1mV EKG signal would be smaller than the common-mode noise. Such a large common-mode voltage would easily justify the expense of the instrumentation amp chip.  (Note: large DC common-mode voltages don’t matter, as the DC-blocking capacitor I used after the instrumentation amp can eliminate them.)

If the AC common-mode voltage is only ±1 mV, then the Blinky EKG could probably work even with very poor common-mode rejection in the instrumentation amp, and building it out of op amps and discrete resistors is feasible.

So today I tried measuring the common-mode signal.  As it turns out, I had built an EKG amplifier on the protoboard that had a split Rgain resistor (two 12kΩ resistors in series).  The node between the 2 resistors should be at (Vp+Vm)/2, the common-mode voltage.  That node was fed to a unity-gain buffer, so I tried measuring the AC common-mode voltage.

I did this with two different tools: my Fluke 8060A multimeter (which measures RMS voltage) and my BitScope Pocket Analyzer, which can measure peak-to-peak voltage.  With the multimeter, I saw about 20–35 mV RMS, and with the BitScope I saw 90–130 mV peak-to-peak (depending on where my arms were relative to the wires to the electrodes).

Trace of common mode noise recorded by the BitScope Pocket Analyzer, at 20mV/division.  Sorry about the black background—the manufacturers are promising a white background as an option in some future release of the software.  Maybe they'll add some labels for the x and y tick marks at the same time!

Trace of common mode noise recorded by the BitScope Pocket Analyzer, at 20mV/division. The fundamental frequency of the noise is 60Hz.
Sorry about the black background—the manufacturers are promising a white background as an option in some future release of the software. Maybe they’ll add some labels for the x and y tick marks at the same time!

If I take the leads off the electrodes and clip them all together, then the common-mode noise drops below the level that the BitScope can measure (its noise floor is about 10mV). The multimeter also has trouble, reporting values from 0.05mV to 1.5mV. If I put a 3.3MΩ resistor between the two signal leads, and connect one of them to the Vref lead, then the multimeter reports about 13–20 mV RMS and the BitScope around 80–90 mV peak-to-peak.

The common-mode problems are definitely coming from the wiring to the electrodes.  With 130mV peak-to-peak common-mode voltage and a common-mode gain of 0.02 or 0.04, we’d have 2.6–5.2 mV of common-mode noise at the output of an instrumentation amp made with discrete resistors and op amps.  If the gain of the amplifier were around 6.6, we’d have a peak-to-peak signal of about 6.6 mV for the EKG signal, and the details of that signal would be buried in common-mode noise.

If we got lucky with the resistor values and they matched better than the specs claim (which is actually fairly likely), then the Blinky EKG would sort-of work with discrete op amps—as long as the 60Hz noise is no worse that what I saw.  Many people are likely to be in electromagnetically noisier locations than my room, so the common-mode noise is likely to be much higher sometimes, and even with lucky matching of the resistors the common-mode signal is likely to swamp the real signal.

Thus I have confirmed that an integrated instrumentation amp is needed for the Blinky EKG, and there is no point trying to design one using discrete resistors and op amps.

About these ads

2 Comments »

  1. […] 3-op-amp designs) that 1% tolerance on the resistors would produce poor common-mode rejection. In Common-mode noise in EKG, I reported measurements of the common-mode noise with a fairly short twisted-pair connection to […]

    Pingback by Some failed designs | Gas station without pumps — 2013 July 10 @ 12:23 | Reply

  2. […] be a problem. I looked at this problem a year ago in 2-op-amp instrumentation amp and Common-mode noise in EKG, and concluded then that common-mode noise would be too large, but I’m tempted to try again, […]

    Pingback by Instrumentation amp from op amps still fails | Gas station without pumps — 2014 June 26 @ 17:02 | Reply


RSS feed for comments on this post. TrackBack URI

Leave a Reply

Fill in your details below or click an icon to log in:

WordPress.com Logo

You are commenting using your WordPress.com account. Log Out / Change )

Twitter picture

You are commenting using your Twitter account. Log Out / Change )

Facebook photo

You are commenting using your Facebook account. Log Out / Change )

Google+ photo

You are commenting using your Google+ account. Log Out / Change )

Connecting to %s

The Rubric Theme. Blog at WordPress.com.

Follow

Get every new post delivered to your Inbox.

Join 268 other followers

%d bloggers like this: